If you want to integrate the OT-2 with other lab equipment, you have several options that range in both capability and difficulty of execution.

Robot System Overview

The core of the OT-2 is a Raspberry Pi 3+ running Alpine Linux in a container. It connects to your computer either via a wifi network or via a USB to Ethernet adapter - either way, it communicates with your computer using networking technology and therefore can be communicated with and controlled using tools designed to talk to Linux computers via the network, like ‘ssh’.
Because the OT-2’s software is in a container, most of the filesystem _will not persist across boots_. In fact, the only two places that will persist across boots are /data and /mnt/usbdrive. All data should be read from and written to files in /data or /mnt/usbdrive.

The OT-2 also runs a webserver. This webserver is what the Opentrons app talks to, and is how protocols are run.

The webserver is implemented as a Python module, and (at least in this version of the software - the exact details may change going forward) therefore can be imported from a Python script running on the robot to control the robot itself. This module is available on GitHub, at https://github.com/Opentrons/opentrons, in the api subdirectory.

An important thing to note about the python code on github is that it will not control the robot from your computer. The only way to control the robot from your computer is through the webserver. The Python code on GitHub is the code that runs on the robot itself, and the reason we refer customers to it is either:

  1. to use as a reference when writing code that interacts with it on the robot or 
  2. to simulate protocols on their computer. 

Summary: The robot runs a Linux system executes Python code to control its hardware and provide a webserver. The only way to control the robot is through the webserver via one of its network interfaces, or by running Python code on a robot. Data can be persistently saved and loaded in /data.

Case Study Example

In the following sections, we’ll consider a case study of integration: a protocol that requires talking to a balance that communicates via serial, some other instrument that communicates over the network using VISA, and an OT-2.

Integrating By Putting The OT2 In Control

Since the robot runs Linux and is connected to the network, and has USB ports available on the Raspberry Pi, the code that controls the lab equipment could run on the OT2 itself. This is the simplest-to-implement method of integration.
In our case study, you could therefore

  • Connect the balance via a USB/serial adapter (FTDI) to a free USB port on the OT-2’s Raspberry PI
  • Connect the OT-2 and the VISA instrument to the same network
  • Using pip, install pyvisa on the OT-2
  • Write Python code that runs the protocol and runs on the OT-2

The Python code would, in addition to containing the normal protocol commands, connect to the balance via serial and the other instrument using pyVISA. You would run this code by transferring it to the robot using scp:

scp ./my-protocol.py robot-ip:/data/my-protocol.py

and then use ssh to remote into the robot and start the protocol:

ssh root@robot-ip -c "nohup python /data/my_protocol.py"

 (the nohup command means the command will continue even if the SSH session goes away).

We’re transferring the protocol to the robot and executing it directly rather than using the Opentrons app to avoid having to add a bunch of checks if the robot is simulating.
The protocol could save data to files in /data, which you can retrieve after the protocol finishes using scp. It could also, since it is network-connected, transfer the files directly via the network to your computer.

Summary: Connect everything to the OT-2 and write a protocol that controls everything, saving and loading data either to the /data directory on the OT-2 or, via the network, on your computer.

Integrating While Keeping Your Computer In Control: Using Normal Protocols


Another option is to write a Python protocol that connects back to your computer. In our case study, you would:

  • Connect the balance to your computer
  • Have the code that interacts with the VISA instrument on your computer
  • On your computer, run code that listens for the OT-2 protocol to reach out and talk to it
  • Write a Python protocol that is uploaded through the Opentrons app and reaches out and talks to your server

This can be complex, but it does mean that your computer stays in control and you have minimal custom code to write. Consider writing a simple Python server using Flask (http://flask.pocoo.org/), Bottle (https://bottlepy.org/docs/dev/), or AIOHTTP (https://aiohttp.readthedocs.io/en/stable/) (it’s your choice - it will be on your computer. We like AIOHTTP, though).

You set up a route called /update, which something can POST to with a step name, and will return either "wait" or "proceed":
On your computer: This is written with AIOHTTP in mind

from aiohttp import web # You can install aiohttp with pip
import json

async def update(request):
    """
    This function serves POST /update.

    The request should have a json body with a "step" key that at some point
    has the value "done-aspirating".

    It will return a json message with appropriate HTTP status.
    """
    try:
        body = await request.json()
    except json.JSONDecodeError:
        text = await body.text()
        print(f"Request was not json: {text}")
        return web.json_response(status=400, # Bad Request
                                 data={'error': 'bad-request'})
    if 'step' not in body:
        print(f"Body did not have a 'step' key")
        return web.json_response(status=400, # Bad Request
                                 data={'error': 'no-step'})
    if body['step'] == 'done-aspirating':
       # Here you might for instance check a balance
       # attached to the computer to validate apsiration
       print("Robot is done aspirating")
       return web.json_response(status=200, # OK
                                data={'done': True})

# Create and run the actual server application
app = web.Application()
# Install the update function to serve the /update endpoint for POST
app.router.add_post('/update', update)
# Run the application
web.run_app(app, # our application
            host='0.0.0.0', # listen on all network interfaces
                            # (change to 127.0.0.1 to listen only to
                            # requests from this computer for testing)
            port=80)        # the standard http port; may need to
                            # change to something else if another
                            # server is running



The Python protocol gets to a certain step, hits that endpoint, and either waits or proceeds:

import requests # you can install this with pip
import opentrons

pipette = opentrons.instruments.P300_Single('right')
tiprack = opentrons.labware.load('opentrons-tiprack-300ul', '1')
trough = opentrons.labware.load('trough-12row', '2')
pipette.pick_up_tip(tiprack)
pipette.aspirate(10, trough.wells()['A1'])
done = False
# Poll to see if the server wants you to proceed
while not done:
    r = requests.post('http://127.0.0.1/update', json={'step': 'done-aspirating'})
    done = r.json()['done']
pipette.dispense()


This protocol can be simulated with opentrons_simulate while the server runs in another terminal. You will see the server print "Robot is done aspirating".

When this protocol is run through the opentrons run app, it will be simulated before it actually runs. That means that the protocol will run without the robot moving, including making the POST request to your computer. To avoid this, you can only do the request if opentrons.robot.is_simulating()  is False .
If you want to avoid simulation entirely, you can transfer the protocol to the robot with scp and run it directly there, as above.

When the protocol is run on the robot, you must also change 127.0.0.1  to the IP address of your computer.

Summary: Write a protocol that connects to a webserver on your computer. Your computer is connected to the other equipment and waits for the protocol to connect and say it is ready.

Conclusion


One of these two methods should help out most integration tasks. If they don’t, keep in mind that fundamentally the OT-2 is just a computer that runs Linux that happens to be connected to some hardware, and has a Python module for controlling that hardware already installed.

Opentrons has plans to make this sort of task easier in the future; but right now, this is what we’ve got.