Deck calibration?

Deck calibration should be completed once immediately after unboxing your OT-2. 

Following the first deck calibration after receiving the robot, the deck only needs to be recalibrated once or twice per year for maintenance, or if you're troubleshooting calibration issues. 

Pipette calibration?

Pipette calibration is prompted by the Opentrons App each time a protocol is uploaded at time of writing (version 3.15.2 of robot server), but you do not need to complete pipette calibration each time a protocol is uploaded.

Pipette calibration only needs to be redone if:

  1. The pipette has not yet been previously calibrated
  2. The pipette has been unmounted and remounted since the last time it was used in a protocol
  3. The pipette is being used with different tips from those used in its previous run
  4. The deck has been calibrated since the last run

If none of the above circumstances are true, you can skip to labware calibration by clicking your tipracks in the calibration screen, or skip to the run by clicking the "Run" tab of the app.

Labware calibration?

Labware calibration is prompted after pipette calibration has been completed. In all cases, you must complete labware calibration if you have completed pipette calibration. However, if you skip pipette calibration, you may still need to complete labware calibration. 

Labware calibration needs to be completed if:

  1. Pipette calibration has been completed
  2. If there is any labware on the deck being used for the first time
  3. If you are troubleshooting calibration issues

If you're running a previously calibrated protocol with the exact same labware and pipette setup, you don't need to redo labware calibration. If you have handled or repositioned labware, you should do labware calibration before running a protocol as a best practice. If you skip calibrating labware, your OT-2 will use data from the most recent calibration.